Abstract:
This document gives a summary analysis for the latent behaviors on MIPS
24 um array. The
analysis intends to characterize the latency due to a bright stim at
different background levels.
In general, the latents decay quickly after the stim turn off. The
residual count rate at the
first time sample (~5 sec) is about 1~2% of the stim flash. The decay
curve is best fitted with
a two-exponential function (fast decay + slow decay) with fast decay
constant tau1 and slow
decay constant tau2. Tau1 is in the range of 1.0+/-0.2 seconds and
tau2 is 12+/-5 seconds.
The decay constants and the amplitude of the fast decay are also a
function of backgrounds.
The higher the background, the lower the amplitude of the fast decay
and the longer tau1
(increasing with background), but shorter tau2 (decreasing with background).
The results
also show that the more number of resets does lower the slow decay
constant so that the
array suffers less from latency, while the fast decay constant seems
to be independent
of the number of resets. There is a huge pixel-to-pixel variation in
the time constants
(10~30% in tau1 and 20~50% in tau2) across the array. The amplitude
of the fast decay is
pretty steady across the whole array (varys only a few %), while the
amplitude of the slow
decay varies greatly. The latent contribution is significant for shorter
exposures since
the latency is an very function of time. The latent contribution for
each following DCEs
compared with the background variation is briefly discussed in order
to help setting the
threshold value for latent correction.
This is a data analysis report on the latency@mips 24 um array based
on test experiments,
done in Oct. 2001. The previous test results (experiments done in April
2001) show that the
latent is a strong function of bias boosts as illustrated in Figure
1a and 1b. Those experiments
were done with exposing the array to a bright stim flash (first image
in each sequce of images),
then taking more data after shut-off the stim to monitor the latency.
Figure 1a and 1b show
two data sets taken with bias boost only at the 1st following DCE,
and with bias boosts for
each following DCE respectively.
Figure 1a: Bias boost only at the 1st following
DCE
Figure 1b: Bias boost at all the following DCEs
The test experiments done in Oct. 2001 were concentrated on the latent
behavior when
there is no bias boosts in the following DCEs and at exposing at different
level of background.
Figure 1c and 1d display an example of the new test. Similiar to Figure
1a and 1b, the first
image of each sequence of images is when stim was on. The array was
exposure to another
stim (as background), then take the rest of following DCEs. To demonstrate
the latent, the last
DCE taken was subtracted off from the rest of DCEs to take out the
background pattern
contribution (see Figure 1d).
Figure 1c: An example of new data set. Bright
stim was on at the first image, then shut off,
move array to exposure to anther stim as background for the rest of
DCEs.
Figure 1d: Same as 1c, but the following DCEs have been subtracted
the last DCE (background).
Experiments and Data Reduction:
The new experiments were done for different background and both 4 and
10 sec exposures
to test the influence of number of resets. The array was exposed to
bright stim flash (1 DCE),
then shut off to measure the residual counts for other 11 DCEs for
10 sec exposure and other 20
DCEs for 4 sec exposure, with constant exposing
to different background levels. Note
that
these experiments were done when preparing for
Brutus testing, and the photon shutter was
open, which means the external stims are not
in function (The external stims all bounce off the
photon shutter, so with the shutter open, there
is no bounce). As for internal stims, the
commanding is set up so that we can only use
one at a time. Therefore, the bright stim DCE
does not include the background that was used
later for the rest of the following DCEs. There
is one mips sec delay between the data with stim on and the data with
stim shut-off for turning
off the stim and turning on the background. In each background level,
the experiment was run
through seven cycles, and each cycle has a pair of files with the first
one for the stim flash and
the other for the rest of following DCEs. All the data were taken in
the RAW mode. The data
used in this report were listed in Table 1. Data were reduced with
mips_sloper
with
droop and
linearity corrections. The output data are in both slop (DN/sec) and
ramp (DN) format. To
increase the signal-to-noise ratio, the final data used to analysize
the latent were averaged
over the seven cycles for each background level.
Table 1: data numbers used with stim level + background level
|
|
|
10s DCEs | 10s DCEs | 4s DCEs |
Stim 0x2A + 0x1E: 04.49-05.08 | Stim 0x2E + 0x21: 01.35-01.53 | Stim 0x2C + 0x21: 01.41-01.55 |
Stim 0x2A + 0x20: 05.09-05.27 | Stim 0x2E + 0x23: 03.21-03.39 | Stim 0x2C + 0x23: 01.58-02.12 |
Stim 0x2A + 0x22: 05.33-05.52 | Stim 0x2E + 0x25: 03.40-0.359 | Stim 0x2C + 0x25: 02.15-02.30 |
Stim 0x2A + 0x24: 05.53-06.11 | Stim 0x2E + 0x28: 03.01-03.20 | Stim 0x2C + 0x28: 02.31-02.46 |
4s DCEs | ||
Stim 0x2E + 0x21: 04.06-04.20 | ||
Stim 0x2E + 0x23: 04.21-04.36 | ||
Stim 0x2E + 0x25: 04.41-04.55 | ||
Stim 0x2E + 0x28: 04.57-05.11 |
Analysis Method:
There are four different background levels (bak1, 2, 3, and 4 with intensity
in an increase
order), and two stim flash levels: low ~4950 DN/s (Day 284), ~9200
DN/s (Day 289) and
high ~14600 DN/s (Day 285). The latent at 24um is known to follow two
decay curves:
fast decay (less equal than a sec) and slow decay (a few sec), from
previous CTA tests.
In order to better characterize the fast decay for the 10s DCEs, the
first following DCE
is splited into 4 sub-DECs using the output ramp files from
mips_sloper.
Including the
stim flash (t=0) and the rest of 10 DCEs, each cycle has 15 time sample
points to monitor
the decay curve for 10s DCEs.
In order to compensate the illumination pattern of the stim flash, each
pixel value was
normalized to the asscociated stim flash value. The high stim flash
in the 10s DCEs
(20 reads) was saturated after 10 reads, however, the final counts
rate can be determined
by the non-saturated reads by mips_sloper since the data are
in RAW mode. In order to
see the deviation caused by the illumination pattern, two statistical
areas were used to
get average normalized response values: using a box (30x30) near the
center and using
the whole array (128x128). The average values are listed in Table 2.
Fitting Function:
The fitting function is given as:
y = p(0) + p(1).exp(-t/tau1) + p(3).[ 1 - exp(-t/tau2) ]
y: normalized response
t: time elapse in unit of sec
tau1: fast decay time constant in unit of sec
tau2: slow decay time constant in unit of sec
p(0): fitted background level
p(1): fitted amplitude of the fast decay
p(3): fitted amplitude of the slow decay
The IDL fitting package MPFIT was used to do the latency fitting.
The quality of the fit is shown in the Figure 2 where the data points and associated fitted lines are plotted.
Results:
Table 2a: Normalized Response for Day 284
10s DECs. (The 3rd and the last line are in DN/s unit given the
information of the stim and bacground levels before normailzation.)
|
|
|
|
|
||||
|
|
|
|
|
|
|
|
|
0.000 | 4914.05 | 3474.25 | 4938.31 | 3491.16 | 4957.63 | 3504.93 | 4927.35 | 3483.85 |
0.000 | 1.000e+00 | 1.000e+00 | 1.000e+00 | 1.000e+00 | 1.000e+00 | 1.000e+00 | 1.000e+00 | 1.000e+00 |
5.243 | 1.488e-02 | 1.605e-02 | 3.041e-02 | 3.168e-02 | 6.399e-02 | 6.532e-02 | 1.369e-01 | 1.388e-01 |
6.816 | 1.364e-02 | 1.458e-02 | 2.867e-02 | 2.977e-02 | 6.194e-02 | 6.323e-02 | 1.342e-01 | 1.360e-01 |
8.389 | 1.263e-02 | 1.346e-02 | 2.750e-02 | 2.845e-02 | 6.077e-02 | 6.187e-02 | 1.319e-01 | 1.340e-01 |
9.961 | 1.120e-02 | 1.280e-02 | 2.670e-02 | 2.763e-02 | 5.992e-02 | 6.098e-02 | 1.323e-01 | 1.335e-01 |
22.544 | 1.049e-02 | 1.119e-02 | 2.515e-02 | 2.593e-02 | 5.834e-02 | 5.935e-02 | 1.303e-01 | 1.318e-01 |
33.554 | 9.751e-03 | 1.039e-02 | 2.437e-02 | 2.511e-02 | 5.761e-02 | 5.855e-02 | 1.297e-01 | 1.310e-01 |
44.564 | 9.405e-03 | 1.000e-02 | 2.402e-02 | 2.473e-02 | 5.727e-02 | 5.818e-02 | 1.293e-01 | 1.307e-01 |
55.575 | 9.184e-03 | 9.769e-03 | 2.380e-02 | 2.449e-02 | 5.707e-02 | 5.759e-02 | 1.292e-01 | 1.305e-01 |
66.585 | 9.054e-03 | 9.622e-03 | 2.367e-02 | 2.434e-02 | 5.695e-02 | 5.781e-02 | 1.291e-01 | 1.304e-01 |
77.595 | 8.949e-03 | 9.517e-03 | 2.356e-02 | 2.422e-02 | 5.685e-02 | 5.771e-02 | 1.290e-01 | 1.303e-01 |
88.605 | 8.885e-03 | 9.444e-03 | 2.351e-02 | 2.415e-02 | 5.678e-02 | 5.762e-02 | 1.289e-01 | 1.302e-01 |
99.615 | 8.836e-03 | 9.391e-03 | 2.345e-02 | 2.407e-02 | 5.673e-02 | 5.756e-02 | 1.289e-01 | 1.302e-01 |
110.625 | 8.794e-03 | 9.338e-03 | 2.341e-02 | 2.403e-02 | 5.667e-02 | 5.751e-02 | 1.288e-01 | 1.301e-01 |
121.635 | 8.746e-03 | 9.283e-03 | 2.337e-02 | 2.339e-02 | 5.665e-02 | 5.748e-02 | 1.288e-01 | 1.301e-01 |
121.635 | 42.9616 | 32.2384 | 115.074 | 83.5091 | 279.231 | 200.295 | 637.707 | 455.332 |
Fig 2a: Decay curves for Day 284 10s DCEs.
Black dots are the average values over a box with green
fitting line, and red diamonds are the average values over the whole
array with blue fitting line.
The fitted decay constants are given inside the box of each plot.
Table 2b: Normalized Response for Day 285 10s DECs. (The 3rd and the last line are in DN/s unit.)
|
|
|
|
|
||||
|
|
|
|
|
|
|
|
|
0.000 | 14201.3 | 10045.3 | 14338.2 | 10138.7 | 14399.5 | 10181.9 | 14423.7 | 10199.9 |
0.000 | 1.000e+00 | 1.000e+00 | 1.000e+00 | 1.000e+00 | 1.000e+00 | 1.000e+00 | 1.000e+00 | 1.000e+00 |
5.243 | 1.764e-02 | 1.858e-02 | 3.543e-02 | 3.645e-02 | 7.014e-02 | 7.124e-02 | 1.919e-01 | 1.938e-01 |
6.816 | 1.648e-02 | 1.728e-02 | 3.381e-02 | 3.470e-02 | 6.763e-02 | 6.910e-02 | 1.896e-01 | 1.912e-01 |
8.389 | 1.573e-02 | 1.644e-02 | 3.280e-02 | 3.361e-02 | 6.736e-02 | 6.817e-02 | 1.884e-01 | 1.898e-01 |
9.961 | 1.519e-02 | 1.584e-02 | 3.183e-02 | 3.279e-02 | 6.632e-02 | 6.741e-02 | 1.876e-01 | 1.887e-01 |
22.544 | 1.389e-02 | 1.442e-02 | 3.063e-02 | 3.130e-02 | 6.522e-02 | 6.612e-02 | 1.864e-01 | 1.876e-01 |
33.554 | 1.332e-02 | 1.376e-02 | 3.005e-02 | 3.065e-02 | 6.463e-02 | 6.545e-02 | 1.860e-01 | 1.871e-01 |
44.564 | 1.303e-02 | 1.344e-02 | 2.977e-02 | 3.032e-02 | 6.441e-02 | 6.517e-02 | 1.858e-01 | 1.869e-01 |
55.575 | 1.288e-02 | 1.325e-02 | 2.961e-02 | 3.013e-02 | 6.430e-02 | 6.505e-02 | 1.858e-01 | 1.868e-01 |
66.585 | 1.278e-02 | 1.314e-02 | 2.951e-02 | 3.001e-02 | 6.421e-02 | 6.494e-02 | 1.857e-01 | 1.868e-01 |
77.595 | 1.272e-02 | 1.307e-02 | 2.944e-02 | 2.993e-02 | 6.415e-02 | 6.486e-02 | 1.857e-01 | 1.867e-01 |
88.605 | 1.267e-02 | 1.301e-02 | 2.940e-02 | 2.987e-02 | 6.411e-02 | 6.481e-02 | 1.857e-01 | 1.867e-01 |
99.615 | 1.263e-02 | 1.296e-02 | 2.936e-02 | 2.983e-02 | 6.408e-02 | 6.477e-02 | 1.857e-01 | 1.867e-01 |
110.625 | 1.260e-02 | 1.293e-02 | 2.933e-02 | 2.979e-02 | 6.407e-02 | 6.475e-02 | 1.857e-01 | 1.867e-01 |
121.635 | 1.258e-02 | 1.290e-02 | 2.932e-02 | 2.977e-02 | 6.404e-02 | 6.472e-02 | 1.857e-01 | 1.867e-01 |
121.635 | 178.662 | 129.575 | 420.405 | 301.832 | 922.177 | 658.993 | 2678.97 | 1904.62 |
Fig 2b: Decay curves for Day 285 10s DCEs.
Black dots are the average values over a box with green fitting line,
and red diamonds are the average values over the whole array with blue
fitting line. The fitted decay constants
are given inside the box of each plot.
Table 2c: Normalized Response for Day 285 4s DECs. (The 3rd and the last line are in DN/s unit.)
|
|
|
|
|
||||
|
|
|
|
|
|
|
|
|
0.00 | 14678.4 | 10361.7 | 14652.6 | 10343.8 | 14837.9 | 10470.9 | 14930.2 | 10538.0 |
0.00 | 1.000e+00 | 9.998e-01 | 1.000e+00 | 9.998e-01 | 1.000e+00 | 9.998e-01 | 1.000e+00 | 9.998e-01 |
5.243 | 1.761e-02 | 1.934e-02 | 3.447e-02 | 3.639e-02 | 6.948e-02 | 7.186e-02 | 1.875e-01 | 1.915e-01 |
9.961 | 1.477e-02 | 1.596e-02 | 3.140e-02 | 3.277e-02 | 6.509e-02 | 6.710e-02 | 1.827e-01 | 1.862e-01 |
14.680 | 1.380e-02 | 1.477e-02 | 3.042e-02 | 3.158e-02 | 6.400e-02 | 6.578e-02 | 1.817e-01 | 1.850e-01 |
19.400 | 1.334e-02 | 1.418e-02 | 2.998e-02 | 3.101e-02 | 6.355e-02 | 6.520e-02 | 1.814e-01 | 1.845e-01 |
24.117 | 1.308e-02 | 1.384e-02 | 2.972e-02 | 3.068e-02 | 6.329e-02 | 6.487e-02 | 1.811e-01 | 1.842e-01 |
28.836 | 1.290e-02 | 1.360e-02 | 2.955e-02 | 3.046e-02 | 6.314e-02 | 6.465e-02 | 1.810e-01 | 1.840e-01 |
33.554 | 1.279e-02 | 1.345e-02 | 2.944e-02 | 3.030e-02 | 6.304e-02 | 6.451e-02 | 1.810e-01 | 1.839e-01 |
38.273 | 1.269e-02 | 1.332e-02 | 2.935e-02 | 3.019e-02 | 6.294e-02 | 6.439e-02 | 1.809e-01 | 1.838e-01 |
42.992 | 1.263e-02 | 1.324e-02 | 2.929e-02 | 3.010e-02 | 6.289e-02 | 6.430e-02 | 1.809e-01 | 1.838e-01 |
47.710 | 1.257e-02 | 1.316e-02 | 2.925e-02 | 3.003e-02 | 6.284e-02 | 6.424e-02 | 1.808e-01 | 1.837e-01 |
52.429 | 1.253e-02 | 1.311e-02 | 2.920e-02 | 2.997e-02 | 6.279e-02 | 6.418e-02 | 1.808e-01 | 1.837e-01 |
57.147 | 1.248e-02 | 1.305e-02 | 2.916e-02 | 2.992e-02 | 6.277e-02 | 6.414e-02 | 1.808e-01 | 1.837e-01 |
61.886 | 1.246e-02 | 1.302e-02 | 2.912e-02 | 2.987e-02 | 6.274e-02 | 6.411e-02 | 1.808e-01 | 1.836e-01 |
66.584 | 1.244e-02 | 1.298e-02 | 2.911e-02 | 2.983e-02 | 6.273e-02 | 6.408e-02 | 1.808e-01 | 1.836e-01 |
71.303 | 1.241e-02 | 1.294e-02 | 2.908e-02 | 2.981e-02 | 6.271e-02 | 6.405e-02 | 1.808e-01 | 1.836e-01 |
76.022 | 1.239e-02 | 1.291e-02 | 2.906e-02 | 2.978e-02 | 6.270e-02 | 6.403e-02 | 1.807e-01 | 1.836e-01 |
80.740 | 1.238e-02 | 1.290e-02 | 2.904e-02 | 2.976e-02 | 6.268e-02 | 6.401e-02 | 1.807e-01 | 1.836e-01 |
85.459 | 1.237e-02 | 1.288e-02 | 2.903e-02 | 2.974e-02 | 6.266e-02 | 6.399e-02 | 1.807e-01 | 1.836e-01 |
90.178 | 1.236e-02 | 1.286e-02 | 2.901e-02 | 2.972e-02 | 6.266e-02 | 6.397e-02 | 1.807e-01 | 1.836e-01 |
94.896 | 1.235e-02 | 1.285e-02 | 2.901e-02 | 2.970e-02 | 6.267e-02 | 6.396e-02 | 1.807e-01 | 1.836e-01 |
94.896 | 181.184 | 131.13 | 8 425.038 | 304.271 | 929.624 | 663.798 | 2697.88 | 1920.95 |
Fig 2c: Decay curves for Day 285 4s DCEs.
Black dots are the average values over a box with green fitting line,
and red diamonds are the average values over the whole array with blue
fitting line. The fitted decay constants
are given inside the box of each plot.
Table 2d: Normalized Response for Day 289 4s DECs. (The 3rd and the last line are in DN/s unit.)
|
|
|
|
|
||||
|
|
|
|
|
|
|
|
|
0.000 | 9222.18 | 6112.23 | 9257.12 | 6134.12 | 9307.77 | 6166.42 | 9406.66 | 6231.63 |
0.000 | 1.000e+00 | 1.000e+00 | 1.000e+00 | 1.000e+00 | 9.998e-01 | 9.998e-01 | 9.998e-01 | 9.998e-01 |
5.243 | 2.660e-02 | 5.503e-02 | 1.125e-01 | 3.119e-01 | 2.832e-02 | 5.699e-02 | 1.150e-01 | 3.160e-01 |
9.961 | 2.339e-02 | 5.122e-02 | 1.082e-01 | 3.067e-01 | 2.460e-02 | 5.268e-02 | 1.104e-01 | 3.102e-01 |
14.680 | 2.236e-02 | 5.012e-02 | 1.071e-01 | 3.057e-01 | 2.340e-02 | 5.141e-02 | 1.092e-01 | 3.092e-01 |
19.400 | 2.184e-02 | 4.963e-02 | 1.067e-01 | 3.053e-01 | 2.283e-02 | 5.084e-02 | 1.087e-01 | 3.088e-01 |
24.117 | 2.155e-02 | 4.932e-02 | 1.064e-01 | 3.052e-01 | 2.245e-02 | 5.049e-02 | 1.083e-01 | 3.086e-01 |
28.836 | 2.135e-02 | 4.916e-02 | 1.062e-01 | 3.051e-01 | 2.221e-02 | 5.027e-02 | 1.081e-01 | 3.084e-01 |
33.554 | 2.122e-02 | 4.901e-02 | 1.061e-01 | 3.050e-01 | 2.205e-02 | 5.008e-02 | 1.080e-01 | 3.083e-01 |
38.273 | 2.114e-02 | 4.893e-02 | 1.060e-01 | 3.050e-01 | 2.194e-02 | 4.999e-02 | 1.079e-01 | 3.083e-01 |
42.992 | 2.105e-02 | 4.886e-02 | 1.059e-01 | 3.049e-01 | 2.184e-02 | 4.991e-02 | 1.078e-01 | 3.082e-01 |
47.710 | 2.099e-02 | 4.878e-02 | 1.059e-01 | 3.049e-01 | 2.177e-02 | 4.981e-02 | 1.077e-01 | 3.082e-01 |
52.429 | 2.094e-02 | 4.876e-02 | 1.059e-01 | 3.049e-01 | 2.171e-02 | 4.976e-02 | 1.077e-01 | 3.082e-01 |
57.147 | 2.091e-02 | 4.872e-02 | 1.058e-01 | 3.048e-01 | 2.165e-02 | 4.973e-02 | 1.077e-01 | 3.082e-01 |
61.866 | 2.086e-02 | 4.866e-02 | 1.058e-01 | 3.049e-01 | 2.161e-02 | 4.967e-02 | 1.076e-01 | 3.081e-01 |
66.585 | 2.084e-02 | 4.865e-02 | 1.058e-01 | 3.048e-01 | 2.156e-02 | 4.964e-02 | 1.076e-01 | 3.081e-01 |
71.303 | 2.081e-02 | 4.862e-02 | 1.058e-01 | 3.048e-01 | 2.153e-02 | 4.959e-02 | 1.076e-01 | 3.081e-01 |
76.022 | 2.081e-02 | 4.860e-02 | 1.057e-01 | 3.048e-01 | 2.151e-02 | 4.956e-02 | 1.075e-01 | 3.081e-01 |
80.740 | 2.078e-02 | 4.858e-02 | 1.058e-01 | 3.048e-01 | 2.150e-02 | 4.956e-02 | 1.075e-01 | 3.081e-01 |
85.459 | 2.076e-02 | 4.857e-02 | 1.057e-01 | 3.048e-01 | 2.147e-02 | 4.953e-02 | 1.075e-01 | 3.081e-01 |
90.178 | 2.073e-02 | 4.856e-02 | 1.057e-01 | 3.048e-01 | 2.143e-02 | 4.952e-02 | 1.075e-01 | 3.081e-01 |
94.896 | 2.073e-02 | 4.857e-02 | 1.057e-01 | 3.048e-01 | 2.142e-02 | 4.951e-02 | 1.074e-01 | 3.081e-01 |
94.896 | 191.062 | 129.417 | 449.497 | 301.337 | 983.597 | 658.102 | 2867.16 | 1910.94 |
Fig 2d: Decay curves for Day 289 4s DCEs.
Black dots are the average values over a box with green fitting line,
and red diamonds are the average values over the whole array with blue
fitting line. The fitted decay constants
are given inside the box of each plot.
For this averaging whole array analysis, the latent drops quickly after
the stim turn off.
The count rate of the first time sample (~5 sec) is about 1~2%
of the stim flash.
The tau1 is ~ 1+/- 0.1 sec, and the tau2 is ~ 10 +/- 2 sec for this
analysis. Both the
tau1 and tau2 seem to be a function of background levels. The higher
the background,
the less latent contribution and the short slow decay constant.
Table 3a: Statistics of the fitting parameter for Day 284 - 10s DCEs
|
good pixels | median of p(0) | average of p(0) | sigma of p(0) |
1 | 99.95% | 0.015793 | 0.024006 | 0.270996 (big) |
2 | 99.95% | 0.031667 | 0.039284 | 0.121742 (big) |
3 | 99.97% | 0.066300 | 0.072510 | 0.060157 (big) |
4 | 99.96% | 0.139193 | 0.149617 | 0.214249 (big) |
median of p(1) | average of p(1) | sigma of p(1) | ||
1 | 0.984205 | 0.976170 | 0.270672 (27%) | |
2 | 0.968333 | 0.960896 | 0.121047 (12%) | |
3 | 0.933703 | 0.927669 | 0.058841 (6%) | |
4 | 0.860808 | 0.850578 | 0.214388 (25%) | |
median of p(3) | average of p(3) | sigma of p(3) | ||
1 | -0.006582 | -0.013947 | 0.267467 (big) | |
2 | -0.007460 | -0.014397 | 0.117654 (big) | |
3 | -0.007834 | -0.014003 | 0.057553 (big) | |
4 | -0.007889 | -0.018101 | 0.213220 (big) | |
median of tau1 | average of tau1 | sigma of tau1 | ||
1 | 0.826384 | 0.740214 | 0.264198 (32%) | |
2 | 0.858841 | 0.756456 | 0.276896 (32%) | |
3 | 0.864532 | 0.729355 | 0.309062 (36%) | |
4 | 0.929491 | 0.764076 | 0.343972 (37%) | |
median of tau2 | average of tau2 | sigma of tau2 | ||
1 | 18.3066 | 17.5625 | 7.51056 (41%) | |
2 | 16.9192 | 16.3308 | 7.63495 (45%) | |
3 | 16.0424 | 15.8477 | 9.15577 (57%) | |
4 | 14.9267 | 15.7193 | 10.8409 (72%) |
Figure 3a: Histograms for the fitting parameters in Day 284-10s DCEs
![]() |
![]() |
![]() |
![]() |
![]() |
Figure 4a: Decay
time images for Day 284-10s DCEs. First four images are for fast
decay and the
last four are for slow decay.
Summary remarks:
Table 3b: Statistics of the fitting parameter
for Day 285 - 10s DCEs
Bak # | good pixels | median of p(0) | average of p(0) | sigma of p(0) |
1 | 99.99% | 0.018377 | 0.020002 | 0.016593 (big) |
2 | 99.99% | 0.036205 | 0.037913 | 0.016378 (big) |
3 | 99.99% | 0.070338 | 0.073164 | 0.081765 (big) |
4 | 99.98% | 0.199049 | 0.204236 | 0.019566 (9.8%) |
median of p(1) | average of p(1) | sigma of p(1) | ||
1 | 0.981623 | 0.980180 | 0.009971 (1%) | |
2 | 0.963796 | 0.962267 | 0.009948 (1%) | |
3 | 0.929664 | 0.927018 | 0.080803 (8%) | |
4 | 0.800968 | 0.795946 | 0.016568 (2%) | |
median of p(3) | average of p(3) | sigma of p(3) | ||
1 | -0.005593 | -0.006268 | 0.0197404 (big) | |
2 | -0.006516 | -0.007230 | 0.0163932 (big) | |
3 | -0.005704 | -0.007252 | 0.0798609 (big) | |
4 | -0.008898 | -0.015808 | 0.0204555 (big) | |
median of tau1 | average of tau1 | sigma of tau1 | ||
1 | 0.808061 | 0.812415 | 0.070245 (8%) | |
2 | 0.855063 | 0.855840 | 0.072593 (8%) | |
3 | 0.884799 | 0.864545 | 0.133980 (15%) | |
4 | 0.909255 | 0.767274 | 0.334563 (36%) | |
median of tau2 | average of tau2 | sigma of tau2 | ||
1 | 16.7539 | 17.1294 | 2.16576 (13%) | |
2 | 15.3634 | 15.7481 | 2.75268 (18%) | |
3 | 15.6839 | 15.8531 | 3.52784 (22%) | |
4 | 10.0612 | 10.1756 | 6.15205 (61%) |
Figure 3b: Histograms for the fitting parameters in Day 285-10s DCEs
![]() |
![]() |
![]() |
![]() |
![]() |
Figure 4b: Decay
time images for Day 285-10s DCEs. First four images are for fast
decay and the
last four are for slow decay.
Summary remarks:
Table 3c: Statistics of the fitting parameter
for Day 285 - 4s DCEs
Bak # | good pixels | median of p(0) | average of p(0) | sigma of p(0) |
1 | 99.99% | 0.018377 | 0.019644 | 0.016009 (87%) |
2 | 99.99% | 0.035410 | 0.036534 | 0.013901 (39%) |
3 | 99.99% | 0.070681 | 0.071915 | 0.013824 (19%) |
4 | 99.99% | 0.191022 | 0.193184 | 0.014317 (7%) |
median of p(1) | average of p(1) | sigma of p(1) | ||
1 | 0.981624 | 0.980601 | 0.004656 (1%) | |
2 | 0.964591 | 0.963651 | 0.004845 (1%) | |
3 | 0.929324 | 0.928269 | 0.005855 (1%) | |
4 | 0.808979 | 0.807001 | 0.009622 (1%) | |
median of p(3) | average of p(3) | sigma of p(3) | ||
1 | -0.005763 | -0.006232 | 0.015890 (big) | |
2 | -0.005899 | -0.006383 | 0.013849 (big) | |
3 | -0.006808 | -0.007515 | 0.014082 (big) | |
4 | -0.007087 | -0.009278 | 0.015635 (big) | |
median of tau1 | average of tau1 | sigma of tau1 | ||
1 | 0.847177 | 0.851158 | 0.064767 (7%) | |
2 | 0.864019 | 0.862588 | 0.079260 (9%) | |
3 | 0.923563 | 0.912409 | 0.099213 (10%) | |
4 | 0.961009 | 0.900537 | 0.209569 (21%) | |
median of tau2 | average of tau2 | sigma of tau2 | ||
1 | 12.3638 | 12.6558 | 2.24748 (18%) | |
2 | 12.1971 | 12.6375 | 2.98247 (24%) | |
3 | 10.9767 | 11.5202 | 3.53750 (32%) | |
4 | 9.4716 | 10.5350 | 5.39418 (57%) |
Figure 3c: Histograms for the fitting parameters in Day 285-4s DCEs
![]() |
![]() |
![]() |
![]() |
![]() |
Figure 4c: Decay
time images for Day 285-4s DCEs. First four images are for fast
decay and the
last four are for slow decay.
Summary remarks:
The background values vary diversely across the array, whick reflects the faint background levels. The amplitude of the fast decay is a function of background while the amplitude of the slow decay is not. Both tau1 and tau2 have two peaks in the histrograms, and seem to be a function of background levels. The deviation of tau2 across the array gets larger as the background increases.
Table 3d: Statistics of the fitting parameter
for Day 289 - 4s DCEs
Bak # | good pixels | median of p(0) | average of p(0) | sigma of p(0) |
1 | 99.99% | 0.027300 | 0.028533 | 0.014642 (53%) |
2 | 99.99% | 0.055867 | 0.057274 | 0.014405 (26%) |
3 | 99.99% | 0.113899 | 0.115689 | 0.015844 (14%) |
4 | 99.99% | 0.314515 | 0.317923 | 0.022804 (7%) |
median of p(1) | average of p(1) | sigma of p(1) | ||
1 | 0.972702 | 0.971650 | 0.006393 (0.65%) | |
2 | 0.944135 | 0.942912 | 0.006731 (0.71%) | |
3 | 0.886103 | 0.884494 | 0.010500 (1.2%) | |
4 | 0.685488 | 0.682258 | 0.021291 (3.1%) | |
median of p(3) | average of p(3) | sigma of p(3) | ||
1 | -0.006114 | -0.006703 | 0.014064 (big) | |
2 | -0.006518 | -0.007401 | 0.014453 (big) | |
3 | -0.006444 | -0.007943 | 0.016438 (big) | |
4 | -0.006067 | -0.009662 | 0.024938 (big) | |
median of tau1 | average of tau1 | sigma of tau1 | ||
1 | 0.866123 | 0.856375 | 0.112025 (13%) | |
2 | 0.899684 | 0.875255 | 0.143850 (16%) | |
3 | 0.925513 | 0.871542 | 0.195289 (21%) | |
4 | 1.020920 | 0.918012 | 0.282719 ((28%) | |
median of tau2 | average of tau2 | sigma of tau2 | ||
1 | 12.6101 | 13.1142 | 3.61906 (28%) | |
2 | 12.0453 | 12.6701 | 4.37224 (36%) | |
3 | 11.6221 | 12.3950 | 5.45756 (47%) | |
4 | 9.5892 | 10.3338 | 6.60788 (69%) |
Figure 3d: Histograms for the fitting parameters in Day 289-4s DCEs
![]() |
![]() |
![]() |
![]() |
![]() |
Figure 4d: Decay
time images for Day 289-4s DCEs. First four images are for fast
decay and the
last four are for slow decay.
Summary remarks: (Similar to Day 285-4s DCEs)
Discussion:
In order to generalize the latent behavior over the whole array, median
values of the fitted parameters
(p(0), tau1 and tau2) are better representations than average values.
These median values are listed
in Table 4 with each associated deviation over the whole array, and
plotted in Figure 5.
Table 4: Summary of the decay constants with deviation over the array v.s. different stim and background levels.
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Within the error bars (deviation over the whole array, represents the
variation from pixel to pixel)
in Figure 5, both fast and slow decay constants are not related to
the stim levels (low or high) for
both 4s and 10s DCEs. (Different color represents different stim level.)
Over all, the fast decay
constant increases as background increases, while the slow decay constant
decreases as
background increases for all 4s and 10s DCEs. However, detailed repeatability
for one single pixel
is needed to confirm that the big variation is really caused by the
various behavior from pixel to pixel,
or something else. (ps. repeatability
study is currently under going)
Figure 5: Decay constants v.s. different Stim/Background
settings. The data points are the median values
of the fitted parameters and the error bars are the standard deviation
over the whole array. The error bars
on the y-axis represent a difference in latent behavior from pixel
to pixel. The plot on the left are for 4s
DCEs and the plot on the right are for 10s DCEs.
![]() |
![]() |
4s DCEs: black dots are day 285 (higher stim ~14,600 DN/s) and red dots are day289 (lower stim ~9,000 DN/s) | 10s DCEs: black dots are day 284 (lower stim ~5,000 DN/s) and red dots are day 285 (higher stim ~14,600 DN/s) |
In terms of the effect of the number of resets, the decay constants
from Day 285 both 4s (more
number of resets) and 10s DCEs (less number of resets) at the same
stim and background levels
are plotted in Figure 6. It appears that the fast decay constant is
indepent of the number of resets,
whille the slow decay constant is a function of the number of resets.
The more number of resets
(like 4s DCEs), the smaller slow decay constant, which is consitent
with what James found for
the pin-hole data.
Figure 6: Decay constants v.s. number of resets.
Under the same stim flash and backround level,
black dots represent the 10s DCEs and red diamonds represent the 4s
DCEs.
Note that the first following DCE taken was delayed for 1 mips sec to
wait for the bright stim
shut-off and background turn-on. The real observation won't have this
delay. Nevertheless,
we still would like to know the latent contribution in the following
DCEs (Latent#) (# means
the #th following DCE) compared with the background. We define this
latent in unit of the
sigma above the background as:
Latent1
=[CR(1st)
- CR(last)] / sqrt(CR(last))
Latent2
=[CR(2nd)
- CR(last)] / sqrt(CR(last))
Latent3
=[CR(3rd)
- CR(last)] / sqrt(CR(last))
Latent4
=[CR(4th)
- CR(last)] / sqrt(CR(last))
Latent5
=[CR(5th)
- CR(last)] / sqrt(CR(last))
with
CR(1st) : counts rate for the 1st following DCE
.....
CR(last): counts rate for the last following DCE
(background)
Latent# was
calculated over the whole array for each background. The median , average
and deviation
values of Latent#
are
listed in Table 5. In general, among these four different backgrounds,
Latent1
is larger for the lower background, but
smaller than the higher background. The following Latent2,
Latent3, Latent4, Latent#5
are getting smaller and smaller.
Table 5: Latent contribution at following DCEs in unit of sigma above background
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Since the latent is a function of decay times, the latent is more server
in short exposures (4s DCEs)
than in long exposures (10s DCEs). For the 4s DCEs at the lowest background,
latent contribution is
still around 2 sigma above the background after four DCEs taken.
The Latent1 v.s. its associated
bak/stim for each pixel is plotted
in Figure 8. Colors represent different
background levels using the same color table as before (red the lowest,
pink the highest). Similar
bacground level (same color) is tested at two stim levels (higher on
the left, and lower on the right).
Figure 8: Latent1 v.s. bak/stim.
![]() |
![]() |
Assuming that the latent contribution below 2 sigma above the background
is acceptable and no
need for correction. For 10s DCEs, when the ratio of background over
stim is larger than ~0.05,
the Latent1 is less than 2 sigma
above the background, which means the threshold (bak/stim)
for latent correction is 0.05. For 4s DCEs, when the ratio of
background over stim is larger than
~0.3, the Latent1 is less than
2 sigma above the background, which means the threshold (bak/stim)
for latent correction is 0.3. The latent at the second following DCE
(Latent2) and the third
following DCE (Latent3) is also
needed to be corrected. Similar plots for Latent2
and
Latent3
are shown in Figure 9. Therefore, the threshold is ~0.05 for Latent2,
and ~0.02 for Latent3.
Figure 9: Latent2 and Latent3 at 4s DCEs v.s. bak/stim.
![]() |
![]() |