MIPS_247a GeGa Stim Response


IOC-MIPS_247a       10180C          
Title:     MIPS 70 and 160 um Detector Response to Stimulators            
                   
Objective:     Demonstrate that the MIPS arrays respond to stimulators being turned on.            
                   
Description:     A series of integrations will be taken in each array with the stimulators both off and on.            
including flow / state diagram     1) Observe for 10 10-second exposures with 70um array. Turn on stimulator and repeat exposure pattern.            
      2) Observe for 10 10-second exposures with 160um array. turn on stimulator and repeat exposure pattern.            
      Whenever stimulators are turned on in this test their voltages and currents should be monitored.            
                   
Place in Schedule:   After aperture has been opened.            
(relative to water shed events)                  
                   
Required Conditions:     MIPS detector aliveness tests must have been performed.            
      Backgrounds must be less than the equivalent of 10,000 photons/sec:            
      70um test at 2.60E-08 W/m^2/Sr or T<12K      
      160um test at 7.60E-09 " orT<6K      
                   
                   
Resources:                  
                   
  duration (best estimate)         40 min      
  real time downlink (Y / N)         No      
  special post event actions (Y / N)         No      
  Generates 40 DCEs.                
Outcome:                  
  description         Confirmation that stimulators can be detected.      
  can proceed in parallel with other activity         Yes      
  must outcome be confirmed before next event / test /activity         No      
  method of confirmation (sensor TLM, data analysis, etc.)         Data analysis and sensor TLM      
  estimate of data turn around if required for confirmation         N/A      
                   
Unique or included in planned AOT?       Unique.          
                   
Contingency Plan:                  
"What if..."   If stimulator output is not detected but telemtry indicates that they turned on, the test will            
      need to be repeated using the redundant stimulators            
                   
IOC Critical     Y            
                   
References:                  
                   
Template last updated:     9/5/99            
      M. Rieke            
                   


Last Updated on 10/7/99
By Marcia Rieke