MIPS_280a Verify pipeline Pht(2)


IOC-MIPS_280a     10870C,10760C,10770C,10530C          
Title:   MIPS Super Resolution of the SVS-13 Source in the HH 7-11 Region            
    Note: This region has very bright IR sources -- this test can investigate how much of an impact there will be in an AOR.          
                 
Objective:   Verify scientific validity of MIPS Photometry pipeline products and ERO            
                 
Description:   Acquire normal photometry observations of an outflow region using the MIPS Photometry AOT            
    in the Large Field mode.            
      3s @ 24 microns (20 cycles), 3s @ 70 and 160 microns (20 cycles each),          
      large field, fine scale @ 70 microns.          
                 
    Can be the same as:   Evaluate fully supported AOTs (MIPS Photometry/SR)        
        ERO: Press release observation        
                 
Place in Schedule:   After aperture door open            
    After CTA/instrument cool down and all other instrument testing is complete.            
    For 160um, the backgrd radiance must be less than 5.4E-13 W/cm2/sr (< 6 - 7 K).            
    After Photometry AOT functionality testing is complete.            
    Can be concurrent with other AOT/instrument tests.            
                 
Required Conditions:   CTA and instrument at nominal operating temperature for 160 micron band (mirror at 6 K).            
    PCS performance: nominal pointing and offsetting            
                 
Resources:                
                 
  duration (best estimate)         1 hr    
  real time downlink (Y / N)         No    
  special post event actions (Y / N)         No    
                 
Outcome:                
  description         Fully calibrated MIPS Photometry data products in three bands.    
  can proceed in parallel with other activity         No - dedicated observing    
  must outcome be confirmed before next event / test /activity         No    
  method of confirmation (sensor TLM, data analysis, etc.)         Data analysis    
  estimate of data turn around if required for confirmation         48 hrs    
                 
Unique or included in planned uplink or downlink/analysis tool           Standard pipeline system and standard data analysis tools required.    
            MIPS Photometry/SR AOT required for uplink.    
                 
Contingency Plan:           Revisit pipeline and/or AOT design depending on the nature of failure.    
"What if..."                
                 
IOC Critical           No    
                 
References:                
                 
Template last updated:           6-Sep-99    
Template written by:           William B. Latter    
                 


Last Updated on 10/7/99
By Marcia Rieke