Campaign I: Gordon (11 Mar 2003) ============================ Summary: This campaign has four main tasks - optimizing the 70/160 micron bias settings, 70 micron wide and narrow field (WF & NF) focal plane surveys, optimizing the 24 micron operating temperature, and comparing scan and photometry flat fields (24um) and illumination corrections (70um). The 70um illumination correction data can also be used as a preliminary test of the 160um scan versus photometry illumination corretions. CAID TITLE TIME TIME AOR/IER COOKBOOK ALLOC USED PRESENT COMPLETE ------------------------------------------------------------------------------------------- MIPS-0905 Startup Activities (Si & Ge) 35 27.13 6.3.0 yes MIPS-0246 70/160 um Optimum Bias Setting 660 676.67 6.3.0 yes MIPS-0911 IOC Darks, 70um 30 32.55 6.2.2 yes MIPS-0918 70um Routine Phot IC 21 24.32 A6.2 yes MIPS-0915 70um Routine Scan IC 13 14.28 A6.2 yes MIPS-0313 70um Routine IC (NF) 19 23.05 A6.2 yes MIPS-0132 70um Focal Plane Survey (WF, Coarse) 126 104.00 6.3.0 no MIPS-0913 Ge Anneal and Recovery 12 6.31 6.5.2 yes MIPS-0345 24um Optimum Operating Temperature 90 71.53 6.5.2/A6.2 yes MIPS-0137 70um Focal Plane Survey (NF, Coarse) 130 103.15 6.3.0 no MIPS-0139 SED Slit Location and Orientation 108 87.77 6.3.0 no MIPS-0910 IOC Darks, 24um 15 17.25 6.2.9 yes MIPS-0917 24um Routine Phot FlatField 10 11.46 A6.2 yes MIPS-0914 24um Routine Scan FlatField 13 14.28 A6.2 yes MIPS-0992 Shut-down Activities (Si & Ge) 19 18.47 6.3.0 yes total 1301 1232.22 (68.78 margin) campaign allocation 1353 Notes: MIPS-132, 137, 139 1. Cookbook is not complete is some sections. MIPS-246 1. IERs need to be uploaded to the mips IOC tracking site. I found them at the IST CVS site. 2. 70um bias checked from 30-50 mV 3. 160um bias checked from 15-35 mV MIPS-345 1. Do we need to have the AOR executed at the start of this task? Couldn't we reorder the task to be: temp change AOR 24um dark 160um dark repeat That would save one AOR execution. MIPS-930 1. Duration of task is only 6.3 min, but 70um array needs 10 min to recover. How is this delay scheduled?