CAID: MIPS-115 | Title: 24um Detector Response to Stims | |
System: MIPS | Principal: dlp | Deputy: jmuzerolle |
Objective: | Demonstrate that the MIPS arrays respond to stimulators being turned on. |
Description: | The CSMM is positioned at the 24um dark position and one of the four 24um stimulator states (flat-low, flat-high, flood-low, flood-high) is turned on for 3 DCEs, and then turned off for an equivalent number of DCEs. The process is repeated with all 4 of the 24um stim states. Housekeeping data regarding the stimulator voltages and/or currents should be monitored. |
Day: 33.0631944444467 | Suggested Order: E200 | ||
Must be preceded by these activities (list by CAID): MIPS-100 | |||
Place in Schedule: After aperture has been opened. | |||
Priority: Critical | |||
Conditions | |||
Maximum telescope temperature: | 33 K | ||
Minimum telescope temperature: | 5.5 K | ||
Should activity be repeated if focus changes? | No Repeat if focus changes by 0 mm | ||
Maximum absolute pointing uncertainty: | 100 arcsec (1 sigma radial) | ||
Maximum precision offset uncertainty: | 100 arcsec (1 sigma radial) | ||
Observatory configuration: | Standard | ||
Go/No Go criteria: | CSMM, CE, and detectors must be operational, instrument and telescope temperatures must be stable and within range. | ||
Duration: 20 minutes
| Real time downlink:
 
No | |
Data Volume (uncompressed): 19.5 Mbits | |||
List Blocks, AORs, SERs, IERs, or unique sequences required (uplink): IER | |||
List Analysis tools required (downlink): Comparison w/ stim brightness measured on ground desirable.Inspection of frames adequate to determine if stims functioned. | |||
Estimate of data turn around: 1 hours | |||
Description: Confirmation that stimulators can be detected. | |||
Can proceed in parallel with other activity: Yes | |||
Must outcome be confirmed before next activity: No | |||
Method of Confirmation: Inspection of frames, comparison w/ test data. Inspection of stim TLM. | |||
Success criteria: Telemetry indicates stims were activated, and light from the stims was unambiguously detected on the 24um array. | |||
Contingency plan: Verify agreement between telemetry and signal on array. If stim was 'on' and no light was detected flag that stim as inoperative. If light was detected and telemetry indicates a failure, track telemetry error. If stim brightness is anomalous, activity may need to be repeated w/ different stim voltage. | |||
Comments | |||
References: | |||
Additional Comments:
Generates 24 DCEs worth of 24um SUR data. Discussion: Required Conditions: MIPS detector aliveness tests must have been performed. Backgrounds reaching the detector at the 24um dark position must be less than 1/2 saturation. Taking 10 as the attenuation achieved at the dark position, this requires the telescope temperature to be at or below 50K. Success Criteria: Telemetry indicates stims were activated, and light from the stims was unambiguously detected on the 24um array. Possible failure modes are: Telemetry indicates any stim was not activated when it should have been. Light from any stim not detected. Stim detected, but telemetry indicates it was not active. Stim detected, but signal is significantly different than expected from ground test. | |||
This activity last updated: 11/14/2001 1:44:25 PM |