CAID: MIPS-192 | Title: Dark Verification, 160 um | |
System: MIPS | Principal: kgordon | Deputy: |
Objective: | Verify that dark current measurements with different exposure times scale. |
Description: | Point to 160 um dark position and take dark current measurements at 3, 4, & 10 seconds and compare them. |
Day: 0 | Suggested Order: | ||
Must be preceded by these activities (list by CAID): | |||
Place in Schedule: | |||
Priority: Necessary | |||
Conditions | |||
Maximum telescope temperature: | 9 K | ||
Minimum telescope temperature: | 5.5 K | ||
Should activity be repeated if focus changes? | No Repeat if focus changes by 0 mm | ||
Maximum absolute pointing uncertainty: | 0 arcsec (1 sigma radial) | ||
Maximum precision offset uncertainty: | 0 arcsec (1 sigma radial) | ||
Observatory configuration: | standard | ||
Go/No Go criteria: | |||
Duration: 15.1 minutes
| Real time downlink:
 
No | |
Data Volume (uncompressed): 168 Mbits | |||
List Blocks, AORs, SERs, IERs, or unique sequences required (uplink): Dark IERs: mips_drk160um_10s35, mips_drk160um_3s101, mips_drk160um_4s77 | |||
List Analysis tools required (downlink): DAT - mips_sloper, mips_caler, mips_enhancer | |||
Estimate of data turn around: 24 hours | |||
Description: Create 160um dark current calibration images. Standard output of mips_enhancer. | |||
Can proceed in parallel with other activity: No | |||
Must outcome be confirmed before next activity: No | |||
Method of Confirmation: Create dark current images. Compare the scale the all the images to 10 second and see if they are equivalent. | |||
Success criteria: Good dark current measurement. No weird behaviors (check ramps, stim latents, etc). | |||
Contingency plan: It's a calibration frame. If there are large differences between exposure times then we will have to start taking darks with all the exposure times routinely. | |||
Comments | |||
References: | |||
Additional Comments:
To be scheduled once during IOC/SV. | |||
This activity last updated: 6/14/2002 10:53:40 AM |