// ENTER CAID, CAMPN, and Analyst WITHIN THE QUOTES IN THE FOLLOWING 2 LINES ...
$caid="900"; // e.g. $caid = "90";
$campn="C1"; // e.g. $campn = "D2" or $campn = "Q";
$analyst="D. Kelly"; // e.g. $analyst="D. Kelly";
// YOU CAN SKIP DOWN TO THE OUTCOME SECTION NOW.
include ("IOC_connect.php"); // this line isn't for you
include ("IOC_log_fns.php"); // this line isn't for you
$db = ioc_db_connect();
$query="select * from ioc_tasks where camp_crnt = '$campn' and caid = $caid limit 1";
$result = mysql_query($query);
if (mysql_num_rows($result) != 1) {
echo "
Contact Stansberry - ";
echo " there is a problem w/ your write-up.
\n";
return ;
}
$row = mysql_fetch_array($result);
$title = $row["title"];
$princ = $row["principal"];
$deputy= $row["deputy"];
$campn0 = $row["campn0"];
$aorkeys = $row["aorkeys"];
// get real name of principal, deputies
$princ = ioc_get_person($princ);
$princ = $princ[0];
$deps = explode(",",$deputy);
foreach ($deps as $depty) {
$depty = trim($depty);
$depty = ioc_get_person($depty);
$depty = $depty[0];
$depty = explode(",",$depty);
$depty = $depty[0]; // last names only
$deplist[] = $depty;
}
$deplist = implode(", ", $deplist);
$caid = sprintf("%03d",$caid);
$file = "mips-".$caid.$campn.".analysis.php";
// END PHP.
?>
echo "MIPS-$caid, Campaign $campn IOC/SV Analysis"; ?>
if (file_exists("master_header.php")) { include ("master_header.php");} ?>
echo "MIPS-$caid, Campaign $campn IOC/SV Analysis"; ?>
echo "$title \n"; ?>
Principal:
Deputy:
Analyst:
AORKEYS: if (! $aorkeys) {echo "unknown";} else { echo $aorkeys;}?>
Last Updated:
if (file_exists($file)) {echo date("D M d Y, H:i:s", filemtime($file) ) ;} ?>
if (file_exists("analysis_header.php")) { include ("analysis_header.php");} ?>
Task Outcome Summary
- DATA STATUS: Nominal
- TASK OUTCOME: Nominal
Abstract
In this task, diagnostic data is collected during a 24um thermal anneal.
The thermal anneal stabilizes the 24um array in preparation for science
operations. The anneal executed properly, and all temperature behavior
was as expected. The temperature plots from this anneal are presented
in the report for metatask MIPS-2050. The instrument temperature was
roughly 0.1-0.15K colder in campaign C than it had been in campaign B,
but it was still 0.06-0.08K warmer than the stabilized temperature
reached by campaign D1. These temperature offsets had very little
effect on the shapes of the temperature curves.
Analysis
We converted the diagnostic data from the thermal anneal into temperatures
and generated temperature plots. These data are compared to the temperature
curves from previous anneals. See the writeup for MIPS-2050 for details.
Results
Tepmerature Plots from Anneal
The temperature plots can be seen in the writeup for MIPS-2050.
The temperature behavior during the anneal was nominal.
Conclusions
The 24um thermal behavior is close to being stable. The thermal conductances
are unchanging, and we are slowly building up a baseline of anneal curves
that we can use for reference to look for anomalies in future anneals.
Output and Deliverable Products
None.
Actions Following Analysis