// ENTER CAID, CAMPN, and Analyst WITHIN THE QUOTES IN THE FOLLOWING 2 LINES ... $caid="917"; // e.g. $caid = "90"; $campn="I"; // e.g. $campn = "D2" or $campn = "Q" or $campn="meta"; $analyst="C. Papovich"; // e.g. $analyst="D. Kelly"; // YOU CAN SKIP DOWN TO THE OUTCOME SECTION NOW. include ("IOC_connect.php"); // this line isn't for you include ("IOC_log_fns.php"); // this line isn't for you $db = ioc_db_connect(); $query="select * from ioc_tasks where camp_crnt = '$campn' and caid = $caid "; $result = mysql_query($query); if (mysql_num_rows($result) < 1) { echo "
Principal:
Deputy:
Analyst:
AORKEYS: if (! $aorkeys) {echo "unknown";} else { echo $aorkeys;}?>
Last Updated: if (file_exists($file)) {echo date("D M d Y, H:i:s", filemtime($file) ) ;} ?>
All images were processed in the same fashion.
We used mips_sloper turning off the row droop correction, with the Dark Images from the appropriate campaigns, then we used mips_caler with the 24micron latent correction and flat-field calibration turned off, and finally mips_enhancer using the same parameter file as in Campaing D1, except that we normalized the input images before doing the median combine. We basically used a 3sigma rejection and MEDIAN combining to construct flatfields.
We constructed a total flat field using all the images combined from both PHOTOMETRY and SCAN map mode at 24um from Campaigns EFGHIJ. The resulting flat field combines 992 DCEs from 10 sky regions spanning Zodi brightnesses from ~ 30 - 51 MJyr/sr (see table below). The AORs, campaign, and sky brightness values are listed in table 1. We also constructed total flat fields using the data only from PHOTOMETRY and SCAN MAP, using 560 and 432 total DCEs, respectively.
Campaign | Region | Zodi Brightness at 24 um [MJy/sr] | AORKEY |
PHOTOMETRY MODE | |||
E | zodi_11b | 44 | 6770176 |
E | zodi_10b | 31 | 6770432 |
F | zodi_11a | 30 | 6941440 |
F | zodi_11c | 31 | 6945536 |
F | zodi_27a | 45 | 6942976 |
F | zodi_27b | 42 | 6943232 |
G | zodi_12a | 51 | 6936832 |
H | zodi_12b | 49 | 7141632 |
H | zodi_28a | 42 | 7141888 |
I | zodi_12a | 45 | 7164928 |
SCAN MODE | |||
I | zodi_13a | 51 | 7164160 |
J | zodi_13a | 51 | 7638528 |
The combined 24 um flatfields using the separate and combined PHOTOMETRY and SCAN-MAP data from Campaigns EFGHIJ is shown in Figure 1. Figure2 shows the corresponding histograms of the DN/s for the three flatfields. Statistics from the images are listed in table 2.
![]() |
![]() |
![]() |
Flatfield Observing Mode | Number of DCEs | Mean | Median | Mode | Std Dev. | Skew |
PHOTOMETRY and SCAN-MAP | 992 | 0.9945 | 1.0000 | 1.011 | 0.0535 | -0.3477 |
PHOTOMETRY | 560 | 0.9943 | 1.0000 | 1.011 | 0.0531 | -0.2781 |
SCAN-MAP | 432 | 0.9947 | 1.0000 | 1.011 | 0.0549 | -0.3352 |
![]() |
![]() |
![]() |
In figure 3, we show the ratio of the total flat field constructed using data from the SCAN-MAP mode to the total flat field constructed using data from PHOTOMETRY mode, and histrogram of pixel values in the ratio. The residuals seen in the ratio are quite low-level: the standard deviation in the ratio is 0.0031, which is comparable to the number-counting uncertainties. Improving rms requires an order of magnitude increase in the number of images over the ~500 used here for both the PHOTOMETRY and SCAN-MAP flat fields. Therefore, the PHOTOMETRY and SCAN-MAP flatfields are consistent with each other at the ~0.3% level. However, to remove the flatfield residuals from image reductions, we need to use flat fields derived for each scan-mirror position (see below). Therefore, this advocates constructing flat fields using both PHOTOMETRY and SCAN-MAP modes, and using the appropriate flatfield for each scan-mirror position angle.
![]() |
![]() |
We constructed scan-mirror-dependent flatfields using the scripts written by K. Misselt (in MakeFlat.pro). Briefly, these scripts read in a list of AORs, and constructs new multi-extention FITS images that contain all DCEs with the same CSM_PRED value. These scripts then apply mips_sloper, mips_caler, and mips_enhancer with the appropriate values (see above) to construct flat fields at each CSM_PRED value. Using this script we constructed scan-mirror dependent flat fields using both the PHOTOMETRY and SCAN-MAP mode data (which are taken with different CSM_PRED values).
In figure 4, we show the 24 um flat fields from the total PHOTOMETRY (averaged over all CSM_PRED values), and three example flat fields from data with the same Scan Mirror position (with CSM_PRED values 1854.50, 1929.00, and 2149.50). Note the image 'artifacts' (or image 'crud') that moves in the y-direction as a function of scan-mirror position (indicated by the red circle in the figure). Low-level features are highly visible in the scan-mirror depenedent flat fields. This is illustrated in figure 5, which shows the ratio of flat fields for data with CSM_PRED values of 1864.50 to data with CSM_PRED value of 2149.50. The 'bright' and 'dark' regions that appear as ying and yang opposites show the flat-field dependent 'crud'.
We have reanalyzed MIPS 24 um observations of the standard star HD159330 (from task mips-920) using the scan-mirror dependent flatfield. Figure 6 shows observations of HD159330 at scan-rmirror position 2106.5 and reduced using both the total flatfield (with no scan-mirror dependency) and using the scan-mirror dependent flatfield. Using the scan-mirror dependent flat field appears to remove all visible image artifacts.
From the 240 DCEs for HD 159330 reduced with the total PHOTOMETRY flatfield, the mean and std dev of the flux at 24 um are: mean=68614.3 DN/s; stddev= 1232.4 DN/s. Similarily, using the flatfields for each individual position angle, the mean and std dev of the 24 um flux are: mean= 68147.2 DN/s; STDDEV= 1279.2 DN/s. The mean photometry is consistent well within the stddev of the photometry. Note the increase in the scatter of photometry when using the individual flat fields in the reduction of each observation. This is likely due to the lower number of DCEs used to construct the flat field at each scan mirror position (e.g., there are 7 scan-mirror position angles in PHOTOMETRY mode, 6 of the positions have 70 DCEs per flatfield, while one has 140; cf. the 560 DCEs used to construct the total flat).
To check photometric accuracy as a function of array position (see for example, the analysis of mips-917 for campaign f), Figure 7 plots the measured photometry as a function of the y-pixel coordinate of the centroid of the standard star HD 159330 for data reduced using the total PHOTOMETRY-mode flat field, and using the scan-mirror-dependent flat field. Note that there seems to be little variation in photometry with x-pixel coordinate. The slight gradient seen in the photometry with y-pixel see in previous analyses persists here in both the data-reductions. The gradient is essentially the same for each the reductions using each flatfield. The change is roughly delta[ log(DN/s) ] / delta(y) = -0.00027 using the total PHOTOMETRY mode flatfield, which mutliplied by 128 pixels corresponds to a change of 0.035 dex in DN/s over the whole array. The change is roughly delta[ log(DN/s) ] / delta(y) = -0.00028 using the total scan-mirror dependent flatfields, which mutliplied by 128 pixels corresponds to a change of 0.035 dex in DN/s over the whole array. Therefore, using the scan-mirror dependent flat fields does not greatly affect the resulting source photometry.
Figure7.Photometry of the standard HD159330 as a function of y-pixel coordinate on the 24um array. The red solid line indicates the mean flux (in log DN/s) and the dashed lines show the 1 sigma standard deviation in both figures. The solid blue line shows the linear-regression fit to the data points with fit in the plot inset. LEFT Data reduction using total PHOTOMETRY-mode flatfields. RIGHT Data reduction using flatfield derived for each scan-mirror position angle.
Conclusions
We have constructed MIPS 24 um flatfields obtained in both PHOTOMETRY and SCAN map modes using data from campains EFGHIJ (992 total DCEs) from ten different regions of the sky. We have constructed both individual flatfields from data taken in each mode using 560 DCEs in PHOTOMETRTY mode and the 432 in SCAN-map mode, as well as a total superflat. The standard deviation of the 24 um Flatfield over the entire array is approximately 5%, but this statistics includes the illumination pattern of the 24 um array (see Analysis writeup of Flat Fields mips-917 for Campaign D2). Comparing photometry of the standard star HD 159330 at various positions across the array, we find that the photometry varies by less than 2.0 percent. Further analysis of the HD159330 observations using data reduced with scan-mirror-dependent flat fields shows no improvement in the photometric accuracy over that using the total flat field, but does effectively remove all flatfielding artifacts, greatly improving the image aesthetics. Furthermore, we find no variation in flatfield as a function of sky background brightness (for sky brightnesses in the range of ~ 30-50 Myr/sr). Variation in photometry persists at the 2.0% level along the y-pixel coordinate values, which probably corresponds to known problems from the 2nd-read in the onboard 24 um processing.
Output and Deliverable Products
MIPS 24 um Flatfields to be used to reduce data obtained in subsequent campaigns.
Actions Following Analysis